• LinkedIn
  • Facebook
  • About
  • News
  • Contact
  • 0Shopping Cart
Chivaune Technologies
  • Biomedical Test Equipment
      • Cardiac Output Simulators
      • Defibrillator and Pacemaker Analysers
      • ECG Simulators
      • EEG Simulators
      • Electrical Safety Analysers
      • Endoscope Testing Solutions
      • ESU Analysers
      • Flow Analysers
      • Foetal Heart Simulators
      • Incubator Testing
      • Infusion Device Analyser
      • Multi-Parameter Patient Simulators
      • NIBP Simulators
      • Patient Simulator Bundles / Kits
      • Pressure and Vacuum Meters
      • Sp02 Simulators
      • Tympanic Thermometer Tester
      • Ventilator Testing Solutions
      • Ventilator Test Lungs
  • Calibration
  • AV Solutions
    • SIMStation
  • Catalogues
  • Enquire Today
  • Search
  • Menu Menu
Delta 3300 Defibrillator / Transcutaneous Pacemaker Analyser - Netech Biomedical Instruments - Chivaune Technologies

Delta 3300 Defibrillator / Transcutaneous Pacemaker Analyser

Defibrillator / Transcutaneous Pacemaker Analyser

The DELTA 3300 Defibrillator / Transcutaneous Pacemaker Analyser is a precision instrument for testing defibrillators and external (Transcutaneous) Pacemakers. It combines the complete testing functions of a Defibrillator Tester, Transcutaneous Pacemaker Analyser, and 12-Lead ECG / Arrhythmia Simulator into one compact, lightweight, and easy-to-use instrument.

The Delta 3300 tests all Defibrillators including AEDs. It measures both monophasic and Bi-Phasic waveforms. It is compact, lightweight and easy to use. The test results can be saved in the internal memory and uploaded to the PC and printed using ProComPrint software (CD included with the unit).

Enquire Now
Back to Products
  • Description
  • Additional information

Description

PRODUCT HIGHLIGHTS

  • Energy and cardioversion measurement
  • Peak voltage and current reading
  • Waveform Storage and Playback
  • Tests all Automated defibrillators
  • Pacemaker loads from 50 to 1100 Ohm
  • Pacer Amplitude, Rate, Pulse width
  • Demand Sensitivity, Refractory period, immunity and pulse Energy
  • 12 Lead ECG, Arrhythmia, Performance
  • Li-Ion Rechargeable Battery

FEATURES

  • Large LCD Display with Backlight
  • Simple and easy to use
  • Compact & Lightweight (3lbs)
  • Test results can be saved & uploaded to PC using Netech ProComPrint software
  • Li-Ion Rechargeable Battery
  • Hard Carrying case
  • 2 Year Warranty
  • Incredible Value

Additional information

Brand

Netech

Specifications

DEFIBRILLATOR ANALYZER

Energy Output Measurement:
Maximum Energy: 800 J
Display Resolution: Based on Auto Range
Auto Range: 2 – 9.99 J, 10-99.9 J, 100-800 J
Resolution: 0.01 J ( 2 -9.99), 0.1 J (10-99.9), 1 J (100-500)
Accuracy: ±1 % of reading ± 0.1 J (2-9.9), ± 1J (10-99), ± 2 % (100-500) ± 1J
Load Resistance: 50 ohms ±1 %; non-inductive (< 1 μH)
Sampling Time: 20 uS
Measurement Time Window: 60 ms
Absolute Max Peak Voltage: 5000 V
Absolute Max Peak Current: 100 A
Pulse Width: 60 ms
High-Level Output:

High Range: 100:1 amplitude attenuation
Low Range: 200:1 amplitude attenuation
Waveform Storage / Playback: Discharge can be viewed via ECG outputs and paddles

Sync Time Measurements:
Timing Window: Starts: 40 ms at each R wave peak
Test Waveforms: All Waveform simulations available
Delay Time Accuracy: ±1 ms

Charge Time Measurement: 30 to 60 in steps of 15 seconds

TRANSCUTANEOUS PACEMAKER ANALYZER
Test Load Range: 50 -1100 Ohms in steps of 100 Ohms.
Accuracy: 50 – 500 ohm ±1 %,  500 – 1100 ohm ±1.5 %

Pulse Measurements:
Amplitude: 4 to 400 mA (50 ohm load)
Accuracy: ±1 % or ± 1 mA
Max. amplitude: 400 mA (50 Ohm Load)
Rate: 30 to 800 ppm
Accuracy: ±1 % or 2 ppm
Pulse Width: 0.5 to 50 ms
Accuracy: ±1 % or ±0.3 ms

Demand Sensitivity Test:
Waveforms: QRS, Square(SQR), Triangle(TRI), and Haversine (SSQ)
Waveform width: 40, 50, 60, 70, 80, 90, 100, 200, 300, 400 ms
Pacer Rate: 30 to 120 ppm

ECG Output:
Amplitude: User Selectable 0.1 – 5 mV

Pacer Input (Load dependent):
Amplitude (50 ohm): 0 – 10 mV
Resolution (50 ohm): 25 μV
Amplitude: (500 ohm): 0 – 100 mV
Resolution: (500 ohm): 1 mV
Accuracy: ±2 %

Immunity Test:
Interference Signal: 50/60 Hz
ECG signal Amplitude: 0.5 to 5.0 mV
Refractory Period Measurement: 20 to 500 ms (both Pacing and Sensing)
Accuracy: ±5 ms

PATIENT SIMULATOR

ECG General:
Lead Configuration: 12-lead simulation RL, RA, LA, LL, V1-6
Output Impedance: RL to All leads 1000 Ohms, V Leads: 1000 ohms to RL
High Level Output: 0.5 V/mV of low level (lead II)
Rates: 30, 40, 50, 60, 70, 80, 90, 100, 110, 120,130, 140, 150, 160, 170, 180, 190, 200, 220, 240, 260, 280, 300 & 310 BPM
Accuracy: ±1 %
Amplitude: User Selectable 0.1 – 5 mV (Lead II)
Accuracy: ±2 % @ Lead II (SQR)
Arrhythmia Selections: Ventricular Fibrillation Coarse (VFBC)

Ventricular Fibrillation Fine (VFBF)

Ventricular Tachycardia (VTAC)

Atrial Fibrillation (AFIB)
Performance Waveforms: Sine, Square, Triangle
Rate: 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 Hz.
Amplitude: 0.1, 0.25, 0.50, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mV.
Amplitude accuracy: ±2 % (Square Wave -Lead II 1.0 mV )

Pacer Pulse:
Pulse Width: 10mS
Amplitude: 1.0 mV (Lead II)
Accuracy: ± 2 %

GENERAL INFORMATION

Temperature Range:
Operating: +15°C to +35°C
Storage: 0°C to +50°C

Display:
Type LCD with LED backlight
Data Input/Outputs : RS-232C (1) for Computer control
Power: 7.4 Volt Li-Ion Battery 100-230 VAC input, 850 mA DC Charger
Housing: High impact ABS Plastic case
Dimensions (D x W x H): 10.25 x 6.25 x 2.5 Inch, 26 X 16 x 6.4 Cm
Weight: 1.5 kg / 3.3 lbs

Related products

  • Digimano 2000 Digital Pressure and Vacuum Gauge

    Digimano 2000 Digital Pressure and Vacuum Gauge

    Show Details
  • EndobenchXT™ Portable Scope Tester

    EndobenchXT™ Portable Scope Tester

    Show Details
  • Fermitron Tympanic Thermometer Tester

    Tympanic Thermometer Tester

    Show Details
  • Datrend Phase 3 Defibrillator / Pacer Analyser - Chivaune Technologies Australia and New Zealand

    Phase 3 Defibrillator / Pacer Analyser

    Show Details

  • Chivaune Technologies Pty Ltd
  • +617 3157 8684
  • sales@chivaune.com.au

What We Do

Biomedical Test Equipment
Test Equipment Calibration
AV Solutions

Quick Links

Home
About Us
News
Contact Us

© Copyright 2026 - Chivaune Technologies
  • LinkedIn
  • Facebook
EXPMT 2000 External Pacemaker Analyser Netech EXPMT 2000 External Pacemaker Analyser EndobenchXTB™ Image Quality Tester (Bench-mount model) EndobenchXTB™ Image Quality Tester (Bench-mount model)
Scroll to top

"*" indicates required fields

Limited Online Selection – Full Range Available!


While we currently have a select range of products available on our website, we have access to the full catalogue from our trusted suppliers. If you’re looking for something specific that you don’t see here, complete your details below for access to our supplier catalogues.

How to Order:

    1. Contact Us: Use our Contact Page or give us a call at +617 3157 8684.
    2. Provide Details: Let us know what you’re looking for, and we’ll check availability. If you're not sure which specific product you need - we can help guide you.
    3. Order Processing: Once we confirm availability, we’ll process your order and arrange delivery.
We’re happy to help you find exactly what you need.
Name*
Brands of interest *
Consent
Product Enquiry

  • Enquiry for: Delta 3300 Defibrillator / Transcutaneous Pacemaker Analyser