Specifications |
ECG Rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. Accuracy 0.5%
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is 0.6 X Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as Lead II
High Level: 500 X low level output on Lead II
Accuracy: 2% (1-5mV)
Performance Waverforms (PERF)
Sine, Square, Triangle, Pulse
Frequencies: 0.1 to 0.9 in 0.1 Hz steps
1.0 to 9.0 in 1.0 Hz steps
10 to 100 in 10 Hz steps
Accuracy: 1%
Amplitudes: 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II
Pulse: 20 ms pulse of 1mV amplitude repeated at 4 second intervals
R Wave Detection (RWD)
Widths: 10, 40, 50, 60, 70, 80, 90, 100, 110, and 120 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 mV on Lead II.
Pacer (PCR)
AP: Atrial Pacer set at 70 BPM
ASP: Asynchronous Pacer
QRS Amplitudes: 1 mV
PCR: Pacer pulses alone
Pacer Widths: 0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 ms
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV
VP: Ventricular Pacer set at 70 BPM
AVP: Atrial Ventricular Pacer set at 70 BPM. QRS Amplitude: 1 mV
Pacer Amplitude: -2mV
Respiration
Baseline Impedance: 250, 500, 750, and 1000 Ohms
Delta Impedance: 0.1, 0.5, 1.0, and 1.5 Ohms
Accuracy: 20%
Rates: 15, 30, 60, and 120 BPM
Accuracy: 1%
Apnea: Off, continuous, 12 seconds and 32 seconds
Lead Configuration: Leads I and II
Blood Pressure
Impedance: 350 Ohms
Excitation: 2 to 16 Volts
Sensitivity: 5mV/V/mmHg
Static: 0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, and 300 mmHg
Dynamic: 50/10, 60/20, 70/30, 80/40, 100/60, and 120/80 mmHg. Dynamic waveforms track all NSR rates
Temperature
Compatibility: YSI 400 and 700 Series.
Temperature: 25° and 37° C (77° and 98.6° F)
Accuracy: 2% of setting
General:
Power Requirements: One 9 Volt alkaline battery or optional AC adapter
Temperature Requirements: Operating: 15° to 35° C (59° to 95° F); Storage: 0° to 55C (32° to 131° F)
Dimensions: 5.5"x 3.5" x 1.5" (13.9 x 8.9 x 3.8 cm)
Dimensions: 5.5"x 3.5" x 1.5" (13.9 x 8.9 x 3.8 cm)
|