| Specifications |
DEFIBRILLATOR ANALYZER
Energy Output Measurement:
Maximum Energy: 800 J
Display Resolution: Based on Auto Range
Auto Range: 2 – 9.99 J, 10-99.9 J, 100-800 J
Resolution: 0.01 J ( 2 -9.99), 0.1 J (10-99.9), 1 J (100-500)
Accuracy: ±1 % of reading ± 0.1 J (2-9.9), ± 1J (10-99), ± 2 % (100-500) ± 1J
Load Resistance: 50 ohms ±1 %; non-inductive (< 1 μH)
Sampling Time: 20 uS
Measurement Time Window: 60 ms
Absolute Max Peak Voltage: 5000 V
Absolute Max Peak Current: 100 A
Pulse Width: 60 ms
High-Level Output:
High Range: 100:1 amplitude attenuation
Low Range: 200:1 amplitude attenuation
Waveform Storage / Playback: Discharge can be viewed via ECG outputs and paddles
Sync Time Measurements:
Timing Window: Starts: 40 ms at each R wave peak
Test Waveforms: All Waveform simulations available
Delay Time Accuracy: ±1 ms
Charge Time Measurement: 30 to 60 in steps of 15 seconds
TRANSCUTANEOUS PACEMAKER ANALYZER
Test Load Range: 50 -1100 Ohms in steps of 100 Ohms.
Accuracy: 50 – 500 ohm ±1 %, 500 – 1100 ohm ±1.5 %
Pulse Measurements:
Amplitude: 4 to 400 mA (50 ohm load)
Accuracy: ±1 % or ± 1 mA
Max. amplitude: 400 mA (50 Ohm Load)
Rate: 30 to 800 ppm
Accuracy: ±1 % or 2 ppm
Pulse Width: 0.5 to 50 ms
Accuracy: ±1 % or ±0.3 ms
Demand Sensitivity Test:
Waveforms: QRS, Square(SQR), Triangle(TRI), and Haversine (SSQ)
Waveform width: 40, 50, 60, 70, 80, 90, 100, 200, 300, 400 ms
Pacer Rate: 30 to 120 ppm
ECG Output:
Amplitude: User Selectable 0.1 – 5 mV
Pacer Input (Load dependent):
Amplitude (50 ohm): 0 – 10 mV
Resolution (50 ohm): 25 μV
Amplitude: (500 ohm): 0 – 100 mV
Resolution: (500 ohm): 1 mV
Accuracy: ±2 %
Immunity Test:
Interference Signal: 50/60 Hz
ECG signal Amplitude: 0.5 to 5.0 mV
Refractory Period Measurement: 20 to 500 ms (both Pacing and Sensing)
Accuracy: ±5 ms
PATIENT SIMULATOR
ECG General:
Lead Configuration: 12-lead simulation RL, RA, LA, LL, V1-6
Output Impedance: RL to All leads 1000 Ohms, V Leads: 1000 ohms to RL
High Level Output: 0.5 V/mV of low level (lead II)
Rates: 30, 40, 50, 60, 70, 80, 90, 100, 110, 120,130, 140, 150, 160, 170, 180, 190, 200, 220, 240, 260, 280, 300 & 310 BPM
Accuracy: ±1 %
Amplitude: User Selectable 0.1 – 5 mV (Lead II)
Accuracy: ±2 % @ Lead II (SQR)
Arrhythmia Selections: Ventricular Fibrillation Coarse (VFBC)
Ventricular Fibrillation Fine (VFBF)
Ventricular Tachycardia (VTAC)
Atrial Fibrillation (AFIB)
Performance Waveforms: Sine, Square, Triangle
Rate: 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 Hz.
Amplitude: 0.1, 0.25, 0.50, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mV.
Amplitude accuracy: ±2 % (Square Wave -Lead II 1.0 mV )
Pacer Pulse:
Pulse Width: 10mS
Amplitude: 1.0 mV (Lead II)
Accuracy: ± 2 %
GENERAL INFORMATION
Temperature Range:
Operating: +15°C to +35°C
Storage: 0°C to +50°C
Display:
Type LCD with LED backlight
Data Input/Outputs : RS-232C (1) for Computer control
Power: 7.4 Volt Li-Ion Battery 100-230 VAC input, 850 mA DC Charger
Housing: High impact ABS Plastic case
Dimensions (D x W x H): 10.25 x 6.25 x 2.5 Inch, 26 X 16 x 6.4 Cm
Weight: 1.5 kg / 3.3 lbs
|